Digital Mammography Comparison Chart

Company NameFujifilmGE HealthcareGE HealthcareHologic, Inc.Hologic, Inc.Hologic, Inc.IMSPhilipsPlanmedPlanmedSiemens
Product NameAspire CristalleSenographe PristinaSenographe Essential with SenoClaireSelenia Dimensions 3DSelenia Dimensions 2DSelenia Dimensions Avia PackageGiotto Image 3D/3DLMicrodose SIPlanmed Nuance ExcelPlanmed NuanceInspiration
FDA cleared, year2014510k – November 2016; PMA – March 20172014201120082012Yes, 2011201320112011Yes, 2011
Generator typeFujifilm Hi FrequencyHigh frequency single-phase power supplyHigh frequency single-phase power supplyConstant potential, high-frequency inverter type; 7kW max.; fully integrated into gantryConstant potential, high-frequency inverter type; 7kW max.; fully integrated into gantryConstant potential, high-frequency inverter type; 7kW max.; fully integrated into gantryHigh power (8kW) Microprocessor-controlled high frequency generatorMicroprocessor-controlled high-frequency X-ray generatorPlanmed; constant potential high frequencyPlanmed; constant potential high frequencyMicroproccessor -controlled high frequency generator
kV range22 to 49 in 1 kV increments22 to 49 kV22 to 49 kV20-49 kV in 0.5 kV increments20-49 kV in 0.5 kV increments (limited to 39 kV without tomosynthesis option)20-49 kV in 0.5 kV increments (limited to 39 kV without tomosynthesis option)22 kV to 40 kV (adjustable in 0.5 kV increments)20-4020-35 kV, +/- 2 kV20-35 kV, +/- 2 kV23-35 kv
mAs Range2-600 mAs4 to 600 mAs (depending on track, filter and kV)4 to 500 mAs (depending on track, filter and kV)3-5003-5003-5002-600 with increments of 1 mAs4000LF (0,3 mm): 5 – 550 mAs (Toshiba)/5 – 600 mAs (Varian Mo&W); SF (0.1mm): 5 – 310 mAs (Toshiba)/5 – 400 mAs (Varian Mo&W)LF (0,3 mm): 5 – 550 mAs (Toshiba)/5 – 600 mAs (Varian Mo&W); SF (0.1mm): 5 – 310 mAs (Toshiba)/5 – 400 mAs (Varian Mo&W)Mo anode: 2mAs to 500 mAs mode/5 mAs – 500 in AEC mode, W anode: 2mAs – 630mAs mode/5 mAs to 630 mAs in AEC mode
mA range20 – 200 mAMolybdenum target: 100 mA from 25 to 30kV on large focal spot, 40 mA from 25 to 30 kV on small focal spot; Rhodium target: 62 mA from 25 to 30 kV on large focal spot, 35 mA from 25 to 30 kV on small focal spotMolybdenum target: 100 mA from 25 to 30kV on large focal spot, 40 mA from 25 to 30 kV on small focal spot; Rhodium target: 62 mA from 25 to 30 kV on large focal spot, 35 mA from 25 to 30 kV on small focal spot10-200 microns, large focal spot; 10-50 microns, small focal spot10-200 microns, large focal spot; 10-50 microns, small focal spot10-200 microns, large focal spot; 10-50 microns, small focal spotUp to 150 mA for Large Focus, up to 45 mA for small focus50-210LF (0,3 mm): 80-110mA (Mo, Toshiba)/91-120 mA (Varian, Mo&W) SF (0,1 mm): 20-35mA (Toshiba, Mo)/30-42 mA (Varian Mo&W)LF (0,3 mm): 80-110mA (Mo, Toshiba)/91-120 mA (Varian, Mo&W) SF (0,1 mm): 20-35mA (Toshiba, Mo)/30-42 mA (Varian Mo&W)N/S
Time range, sec.0.01 to 10 sec. (61 steps)N/SN/S0.1-2.5 large focal spot; 10-50 small focal spot0.1-2.5 large focal spot; 10-50 small focal spot0.1-2.5 large focal spot; 10-50 small focal spot20 ms to 5 s4-16, depending on breast thickness and glandularityLF (0.3 mm): 0.1 – 5 s SF (0.1 mm): 0.1 – 9.9 sLF (0.3 mm): 0.1 – 5 s SF (0.1 mm): 0.1 – 9.9 sexposure times, automatic: 10 ms to 4 seconds (large focus); 60 ms to 6 s (small focus)
AEC detectorAuto, semi, manualTwo exposure modes: automatic optimization of parameters (AOP) and manual mode availableTwo exposure modes: automatic optimization of parameters (AOP) and manual mode available2-cell, automatic or manual 7-position photocell placement2-cell, automatic or manual 7-position photocell placement2-cell, automatic or manual 7-position photocell placementAnrad LMAMII detector; detector has an AEC multiple esposimete; operation: a) very short pre-exposure to detect the breast density and b) the thickness difference within compressed and non compressed breast to determine kV, filter and starting grid speed; fully microprocessor controlledSmartAEC exposure mode continuously adjusts exposure during image scan, according to feedback from exposure during image scan, according to feedback from exposure sensor48 automatically selected detector areas48 automatically selected detector areasN/S
Parameters controlledTrack, filter, kVp, mAs, AOP or manualTrack, filter, kVp, mAs, AOP or manualManul, auto-time, auto-kV, auto-filterManul, auto-time, auto-kV, auto-filterManual, auto-time, auto-kV, auto-filterkV, mAs, Ag/Rh filterN/SkV, mAs and filtrationkV, mAs and filtrationN/S
Breast Density and Dose Reporting
Density reporting software (list if it is optional or integrated and who is the software vendor)Optional; vendors: 1. Volpara and 2. VuCompOptional; Volpara Density 3D – specifically designed and cleared fo 3D/ Tomo mammography systemsOptional; Volpara Density 3D – specifically designed and cleared fo 3D/ Tomo mammography systemsOptional, Hologic, Inc.Optional, Hologic, Inc.Optional, Hologic, Inc.N/SPhilips Spectral Breast Density Measurement tool (optional)VolparaDensity (optional)VolparaDensity (optional)N/S
Vendor 1: generates patient density scorecards for each study; volumetric numbers include volume of fibroglandular tissue, breast volume and volumetric breast density; vendor 2: identifies and records the radiographically dense areas of a mammogram, and then maps this finding to a calibrated VuCOMP density categoryBreast density score compliant with BI-RADS fifth editionBreast density score compliant with BI-RADS fifth editionVolumetric breast density, volume of fibroglandular tissue, volume of breast, area breast density, BI-RADS density assessment, etc.Volumetric breast density, volume of fibroglandular tissue, volume of breast, area breast density, BI-RADS density assessment, etc.Volumetric breast density, volume of fibroglandular tissue, volume of breast, area breast density, BI-RADS density assessment, etc.N/SThe Philips Spectral Breast Density Measurement uses the spectral data from the photon-counting detector to make a direct measurement of the fibro-glandular content of the breast through spectral material decomposition; the outputs from this measurement are listed below; it is also possible to configure the system to also save the complete spectral data setN/AN/AN/S
Vendor 1: volumetric numbers include volume of fibroglandular tissue, breast volume and volumetric breast density, along with a Volpara Density Grade (VDG) corresponding to the BI-RADS breast density categories; vendor 2: breast area, dense area, percent breast density, density catergory for each caseBreast density score compliant with BI-RADS fifth editionBreast density score compliant with BI-RADS fifth editionConfigurableConfigurableConfigurableN/SVolumetric percentage of glandular tissue; volume of glandular tissue; total breast volume; MicroDose Density Score (correlated to the BI-RADS breast composition score)Patient information, breast density for each breast and average density for both breasts, VDG/ACR breast density category, volume of fibroglandular tissue, volume of breast, average radiation dose/ image, average pressure applied using compression force/contact areaPatient information, breast density for each breast and average density for both breasts, VDG/ACR breast density category, volume of fibroglandular tissue, volume of breast, average radiation dose/ image, average pressure applied using compression force/contact areaN/S
Dose software record doseVendor 1; yes; vendor 2: noDoseWatch provides dose monitoring and analytics that enables you to drive quality care improvement so you can deliver accurate diagnoses in a safe manner for each patientDoseWatch provides dose monitoring and analytics that enables you to drive quality care improvement so you can deliver accurate diagnoses in a safe manner for each patientOptional, Hologic, Inc.Optional, Hologic, Inc.Optional, Hologic, Inc.N/SYes (part of device software, not part of breast density measurement tool)N/AN/AN/S
How is density information stored or transferredVendor 1: DICOM secondary capture image; vendor 2: DICOM mammography structured report or secondary capture image storageStructured report and secondary captureStructured report and secondary captureRadiation dose SR, MPPS, image headersRadiation dose SR, MPPS, image headersRadiation dose SR, MPPS, image headersN/SThe Spectral Breast Density Measurement parameters are stored as part of the DICOM header of the images and also can be displayed in the DICOM structured report; this enables display on the review station together with the acquired images and archivingN/AN/AN/S
Digital DetectorV
Type (CCD, aSi, aSe)aSe exclusive hexagonal close pattern pixelsPatented needle structure CsI scintillator, single piece constructionPatented needle structure CsI scintillator, single piece constructionaSeaSeaSeAmorphous selenium (aSe)Crystalline silicon with no dead pixelsaSeaSeAmporphous selenium (ase)
Spatial resolution10 lp/mm2-D: 7.1 lp/mm; 3-D: 3.5 lp/mm2-D: 7.1 lp/mm; 3-D: 3.5 lp/mm2-D: 7.1 lp/mm; 3-D: 3.5 lp/mm6 lp/mm10 lpmm5.88 lp/mm5.88 lp/mmN/S
Matrix, pixelsN/S2,850 x 2,394 pixels; approx 13 MB per image3,062 x 2,394 pixels; approx 14 MB per imageYesYesYes2,816 x 3,584 (24 x 30 cm), 2,016 x 2,816 (18 x 24 cm)Large: 24 x 26 cm: 4,915 x 5,355 pixel; small: 24 x 12.5 cm: 4,915 x 2,600 pixel; the numbers are a bit larger than expected from 50 micron pixel size and physical image size because the pixel size is defined 20 mm below the patient support; the effective pixel size on the patient support is 49 micronsN/SN/SPixel: 85 um, image matrix: 24 x 30 cm: 2,816 x 3,584, 18 cmx 24 cm: 2,016 x 2,816
Standard size2,850 x 2,3943,062 x 2,3942,560 x 3,328 and 3,328 x 4,096 at 70 microns2,560 x 3,328 and 3,328 x 4,096 at 70 microns2,560 x 3,328 and 3,328 x 4,096 at 70 micronsN/S~50 Mb uncompressed, standard configuration85 x 85 um85 x 85 um24 x 30 cm
High-resolution size28 micron at 1.8 x mag2,850 x 2,3943,062 x 2,3942,560 x 3,328 and 3,328 x 4,096 at 70 microns2,560 x 3,328 and 3,328 x 4,096 at 70 microns2,560 x 3,328 and 3,328 x 4,096 at 70 micronsN/SN/SN/SN/SN/S
Imaging areaN/SN/SN/S18 x 24 cm and 24 x 29 cm18 x 24 cm and 24 x 29 cm18 x 24 cm and 24 x 29 cmN/SN/SN/SN/SN/S
Standard size, cm18 x 24 cm, 24 x 30 cm24 x 29 cm24 x 30.7 cm18 x 24 cm and 24 x 29 cm18 x 24 cm and 24 x 29 cm18 x 24 cm and 24 x 29 cm17, 14 x 23, 94 (18 x 24) or 23, 94 x 30, 46 (24 x 30)240 x 260 mm; studies show that this size is sufficient for 99% of all breasts, with significant improvements in positioning the patient compared to larger detectors; 240 x 126 mm (configurable only with spot compression)23.94 x 30.46 cm171 x 239 cmSmall: 18 x 24 cm, large: 24 x 30 cm
High-resolution size, cm18 x 24 cm, 24 x 30 cm24 x 29 cm24 x 30.7 cm18 x 24 cm and 24 x 29 cm18 x 24 cm and 24 x 29 cm18 x 24 cm and 24 x 29 cm17, 14 x 23, 94 (18 x 24) or 23, 94 x 30, 46 (24 x 30)N/SN/AN/A16 x 20 cm, mag spot: 9 x 9 cm
Additional FeaturesIncludes QC, 1-shot phantom kit, patented comfort compression paddlesDQE at IEC 62220-2-3 equivalent spectrum, at 75_Gy: 70% (+/-3) at 0.5lp/mm and 64% (+/-3) at 2 lp/mm; Bucky front cover thickness: 49 mm; optimized room for positioning due to the Bucky depth: 470 mmIntegrated, exclusive high transmission cellular (HTC) grid; smart paddle system; Fast paddles; automatic collimationIntegrated, exclusive high transmission cellular (HTC) grid; smart paddle system; Fast paddles; automatic collimationIntegrated, exclusive high transmission cellular (HTC) grid; smart paddle system; Fast paddles; automatic collimationN/SN/SN/SN/SN/S
Acquisition Workstation
WorkstationHP Z420 WorkstationQuad core Intel i5 workstationDual core Kontron workstationThree: AWS 8000, AWS 8000 Mobile and AWS 5000Three: AWS 8000, AWS 8000 Mobile and AWS 5000Three: AWS 8000, AWS 8000 Mobile and AWS 5000MultiCore CPU, 24 GB RAM, Windows 7 embedded 64 bit OS, 1 HDD of 1 TbyteIntel Core i5 or betterHP Z420 workstationHP Z420 workstationN/S
Monitor size, inches21.3 in.2 types of display available: 1 MP LCD monitor; 48 cm (19-in.) medical grade; 1,280 x 1,024 pixels (landscape); high luminance – up to 300 Cd/m2; contrast ratio: 2,000:1; viewing angle: 170 degrees; mounted on a rotating arm for in-room access; 3 MP monitor display: high performance color IPS 3 MP monitor; 54 cm (21.2-in.); 2,048 x 1,536 pixels (landscape); brightness: 1,000 Cd/m2; contrast ratio: 1,400:1; viewing angle: 170 degrees; mounted on a rotating arm for in-room access2 types of display available: 1 MP LCD monitor; 48 cm (19-in.) medical grade; 1,280 x 1,024 pixels (landscape); high luminance – up to 300 Cd/m2; contrast ratio: 2,000:1; viewing angle: 170 degrees; mounted on a rotating arm for in-room access; 3 MP monitor display: high performance color IPS 3 MP monitor; 54 cm (21.2-in.); 2,048 x 1,536 pixels (landscape); brightness: 1,000 Cd/m2; contrast ratio: 1,400:1; viewing angle: 170 degrees; mounted on a rotating arm for in-room accessControl monitor: 17 in. color, touch screen on some models; image monitor: three: 21 in. color 2 MP, 21 in. gray 2 MP, 21 in. gray 3 MPControl monitor: 17 in. color, touch screen on some models; image monitor: three: 21 in. color 2 MP, 21 in. gray 2 MP, 21 in. gray 3 MPControl monitor: 17 in. color, touch screen on some models; image monitor: three: 21 in. color 2 MP, 21 in. gray 2 MP, 21 in. gray 3 MP21.3 in. Barco TFT LED 2 Mpixel – ; – very high brightness (250cd/m2) and resolution for medical use; or 21.3 in. Barco TFT LED diagnostic color 3 Mpixel very high brightness (400 cd/m2)Standard monitor 1,280 x 1,024 DICOM monitor 1,280 x 1,024; 3 MP 2,048 x 1,53621 in.21 in.48 cm
Hard disk capacityDisk 1: 330 GB for 2-D images1 internal 250 GB disk for the system, 1TB for image storage1 internal 500 GB, 7,200 RPMAt least 1 TBAt least 1 TBAt least 1 TB70,000 images on HDDUp to 10,000 image acquisitions1 TB1 TB24,7000 images on hard disk (500 GB)
Memory8 GB32 GB4 GBAt least 6 GBAt least 6 GBAt least 6 GB1 TBN/S8 GB8 GB500 GB
Networking capabilitiesPrinter, DMS network4 Gigabit Ethernet port; DVI display and port connector; DICOM 3.0 platform: query/retrieve modality worklist user; modality performed procedure steps; storage provider and user; storage commitment user; query/retrieve user; basic grayscale print user; verification provider; DICOM-compliant CD, DVD-R/-RW and USB data interchange2 Gigabit Ethernet port; DVI display; DICOM 3.0 platform: modality worklist user; modality performed procedure steps; storage provider; storage commitment user; query/retrieve user; basic grayscale print user; verification provider; DICOM-compliant CD-RW data interchange10, 100, 1,000 base-T ethernet10, 100, 1,000 base-T ethernet10, 100, 1,000 base-T ethernetStorage of MG, DX, CR, MR, US, CT, NM, DBT, SC as SCU/SCP, storage commitment/SCU, query retrieve/SCU/SCP, basic print/SCU, modality worklist/SCU, MPPS/SCU verification/ SCU/SCP, DICOM-compliant DVD/CD (multisession)Gigabit copper network category 5 cable; 100 Mbps copper network RJ45 RJ45 connectors no network connection support for external disk with specification: USB 2.0 interface, hard disk based, no separate power supply, minimum storage capacity 160 GBIntel CT x1 PCIe NICIntel CT x1 PCIe NICYes
Gray levels1,024 gray levelsAcquisition dynamic range 14 bitsAcquisition dynamic range 14 bitsAcquire 14 bits, transmit 12 bits, display 8 bitsAcquire 14 bits, transmit 12 bits, display 8 bitsAcquire 14 bits, transmit 12 bits, display 8 bits14 bitN/S1,024 grey levels1,024 grey levels14
DICOM 3.0 compliantYesN/SN/SYesYesYesYesYesYesYesYes
Additional featuresIntelligent AEC (iAEC) factors breast density and implant recognition to calculate optimal dose for each exam; ISC image processing optimizes contrast and IQ; 50 micron image output display; automatic system start up and calibration for efficient workflow and extended system perfomanceConnectivity features: customizable autopush to multiple DICOM databases, autoprint, autodelete based on storage commitment, connectivity to GE service for remote diagnostic capabilityImage preview time <4 sec; can preview 2-D, 3-D projections, 3-D slices, generated 2-D, contrast 2-D images at AWSImage preview time <4 sec; can preview 2-D, 3-D projections, 3-D slices, generated 2-D, contrast 2-D images at AWSImage preview time <4 sec; can preview 2-D, 3-D projections, 3-D slices, generated 2-D, contrast 2-D images at AWSN/SN/SN/SN/SN/S
Review Workstation
WorkstationDell 1,920 x 1,200 wide-screen monitorSenoIris DiagnoseSenoIris DiagnoseSecurView DXSecurView DXSecurView DXRaffaelloIntelliSpace Breast SolutionHP Z420 workstationHP Z420 workstationN/S
Monitor size, inches2 x Barco 5 MP (21.3 in. diagonal)Dual 5 MP 21.3-in. LCD display; single 12 MP 33.6-in. color LCD display; 19-in. TFT color monitorDual 5 MP 21.3-in. LCD display; single 12 MP 33.6-in. color LCD display; 19-in. TFT color monitor21.3 in. (5 MP) or 30 in.(10 MP)21.3 in. (5 MP) or 30 in.(10 MP)21.3 in. (5 MP) or 30 in.(10 MP)2 x 5 MPixel Barco Coronis monitors; 21.3 in. self-calibrating, mono-chrome monitors portrait format with very high resolution and brightness dedicated to digital mammography; Barco QAWEB for calibration and quality control546.4 x 185.4 x 365 493.9 mm (21.51 x 7.30 x 14.37 19.44)2 x 21 in.2 x 21 in.N/S
Hard disk capacity256 GB1 x 256 GB, 2 x 1 TB SSD (additional 1 TB optional)1 x 256 GB, 2 x 1 TB SSD (additional 1 TB optional)At least 1.7 TB, raid 5At least 1.7 TB, raid 5At least 1.7 TB, raid 51 x 1 TB (raid configuration)N/S1 TB1 TBN/S
Memory16 GB32 GB RAM32 GB RAMAt least 6 GBAt least 6 GBAt least 6 GB16 GB RAM8 GB (2 x 4 GB), 1,333 Mhz, RDIMM, low volt, dual rank, x88 GB8 GBN/S
Networking capabilities10/100/1,000 ethernet10/100/1,000 base T Ethernet10/100/1,000 base T Ethernet10, 100, 1,000 base-T ethernet10, 100, 1,000 base-T ethernet10, 100, 1,000 base-T ethernet100/1,000 Mbits/sN/SIntel CT x1 PCIe NICIntel CT x1 PCIe NICN/S
Archive/storageN/SEnterprise archiveEnterprise archiveSecurXchange archiveSecurXchange archiveSecurXchange archiveN/SN/SN/SN/SN/S
TypeImages stored in PACSDepartmental PACSDepartmental PACSDepartmental PACSDepartmental PACSDepartmental PACSCPU with processor Intel Core i5-2500 3.30 GHzN/SN/AN/AN/S
CapacityN/AN/SN/SAt least 3 TBAt least 3 TBAt least 3 TB2 x 1 TB (raid configuration)N/SN/AN/AN/S
Gray levels1,024 gray levelsN/SN/S12 bits12 bits12 bitsN/SN/S1,024 gray levels1,024 gray levelsN/S
DICOM 3.0 compliantYesN/SN/SYesYesYesYesYesYesYesN/S
Additional featuresN/SMeetings requirements of IHE radiology mammography imaging integration profile-imagine display actor; capable of accepting DICOM breast tomosynthesis objects; explore DBT images quickly and easily with the V-Preview navigation tools (Navigate, Inspector, Bookmark)Meetings requirements of IHE radiology mammography imaging integration profile-imagine display actor; capable of accepting DICOM breast tomosynthesis objects; explore DBT images quickly and easily with the V-Preview navigation tools (Navigate, Inspector, Bookmark)Custom workflow keypad; application synchronization; multimodality viewing; barcode scanner; workstation cluster managerCustom workflow keypad; application synchronization; multimodality viewing; barcode scanner; workstation cluster managerCustom workflow keypad; application synchronization; multimodality viewing; barcode scanner; workstation cluster managerN/SMulti-modality, vendor neutralDigiPadDigiPadN/S
X-Ray Tube
Anode typeRotatingDual track anode: molybenum (Mo) enriched with vanadium and rhodium (Rh)Dual track anode: molybenum (Mo) enriched with vanadium and rhodium (Rh)Rotating tungstenRotating tungstenTungsten (W) doped with rhenium (Re)/dual anode angleTungstenRotating bi-angular anodeRotating bi-angular anodeDual
Heat capacity, HU300,000 HUAnode heat storage capacity: 250 kJ (340 kHU)Anode heat storage capacity: 250 kJ (340 kHU)300,000 (222 kJ)300,000 (222 kJ)300000445kJ (600 kHU)Tube housing 700 kHUTube housing 700 kHU1 800 000j 430 000 HU
Heat dissipation rate, HU/min.60,000 HU/min500 W (40 kHU/min)N/S6000060000600000.7 kW (maximum workflow)Anode: 56 kHU/minAnode: 56 kHU/minN/S
Target/filter combinationsW/RHMo/Mo and Rh/AgN/S2-D: W Rh, W Ag; 3-D: W Al2-D: W Rh, W Ag; 3-D: W AlW/Rh; W/Ag0.40 mm (0.016-in.) AlMo/Mo or RH; W/RH or AgMo/Mo or RH; W/RH or AgMO anode: 2mAs to 500 mAs mode/5 mAs – 500 in AEC mode, W anode: 2 mAs -630 mAs mode/5 mAs to 630 mAs in AEC mode
Focal spot size, mm0.1 and 0.30.1 and 0.3 IEC on each target0.1 and 0.3 IEC on each target0.1 and 0.30.1 and 0.30.1/0.30.3 mm (0.012-in.) according to IEC 60,3360.3 mm/0.1 mm0.3 mm/0.1 mmMolybdeum: 0.1/0.3; tungsten: 0.1 /0.3
Positioning Assembly
CollimationN/SFilters: molybenum 0.030 mm; silver 0.030 mmFilters: molybenum 0.030 mm; rhodium 0.025 mmAutomatic or user-selectableAutomatic or user-selectableAutomatic/manualX-ray field is fixedN/S
18 x 24 cmYesYesYesYesYesAutomaticN/SYesYesYes
24 x 30 cmYesYesYesYes (24 x 29 cm)Yes (24 x 29 cm)AutomaticN/SYesNoYes
Movement locksElectromagneticGantry locked when compression force appliedGantry locked when compression force appliedFail-safe electromagnetic locksFail-safe electromagnetic locksYesAll movements are motorizedNo, motorized movementsNo, motorized movementsN/S
Assembly movementN/SN/SN/SYesYesYesThe gantry movement allow both face-to-face patient positioning and conventional (technologist at the side of the patient); the 3-D movement is useful in case of disabled patientN/SN/S
Rotation, degreesnegative 190 to positive 190 degrees180 to – 180 degrees185 to – 165 degrees2-D: 195 to -155; biopsy and 3-D: 180 to -1402-D: 195 to -155; biopsy and 3-D: 180 to -1402-D: 195 to -155; biopsy and 3-D: 180 to -140180 degrees to -180 degrees – isocentric programmable-100 to + 180, isocentricMotorized, -135 degrees to +180 degreesMotorized, -135 degrees to +180 degreesPlus 180 to minus 180
Vertical, cm (inches)69 to 150 (27.2 to 59.1)Floor to image receptor distance: from 65 cm to 150 cmFloor to image receptor distance: from 65 cm to 150 cm71-142 (27.8-56)71-142 (27.8-56)75 cm range (65 cm-140 cm)76-138Motorized, 800 mm to 1,400 mmMotorized, 800 mm to 1,400 mmN/S
SID, cm6566 cm66 cm707065 cm (25.6 in.)66650 mm650 mm65 cm
Scale guideYes, newtons (N) for force and millimeters (mm) for thicknessYesYesYes, two digital displaysYes, two digital displaysYes, two digital displaysYesN/AN/AN/AN/S
HandswitchYes, also dual foot switchesDual foot-pedals for column height and compression adjustmentsDual foot-pedals for column height and compression adjustmentsYes, on both sides of the deviceYes, on both sides of the deviceYesN/SNoNoN/S
Radiation Output
mR/sec. at 28 kVp540 mR/sec with W/Rh tube using 50 micron filterN/SN/S>230>230>23021uGy/mAs: 3.05uG/s at 30 kVN/S2,100 mR/s2,100 mR/sN/S
Radiation Shield
L x W, cm (inches)108 x 60.4cm (42.5 x 24 in.)Choice between two radiation shields: integrated to the control console (height: 190 cm or 220 cm); standaloneChoice between two radiation shields: integrated to the control console (height: 190 cm or 220 cm); standalone203 x 86 (80 x 34)203 x 86 (80 x 34)60 x 193 cm732 x 2874 x 30.5 in.74 x 30.5 in.N/S
Thickness0.5 mm Pb equivalent; designed to meet the relevant requirements of 21 CFR 1,020 and IEC60601-2-45:2001N/SN/S6 mm tempered glass, 0.5 mm leaded glass equivalent6 mm tempered glass, 0.5 mm leaded glass equivalent6 mm tempered glass, 0.5 mm leaded glass equivalent0.5 mm Pb0.5 mm0.5 mmPb led equivalent0.5 mmPb led equivalentN/S
Compression systemAuto compression: 0N to 200 N; manual compression: 0N to 200 NMotorized/manualMotorized/manualYes, with lightweight, flexing paddles, automatically shifting paddle positions and auto-sensing paddle identificationYes, with lightweight, flexing paddles, automatically shifting paddle positions and auto-sensing paddle identificationYes, with lightweight, flexing paddles, automatically shifting paddle positions and auto-sensing paddle identificationSCS-sensitive compression system-double power driven speed: fast up to the breast and very slow when compression starts; foot-pedal controlMotorized compressionMotorizedMotorized3kg to 20 kg automatic and manual adjustments
Force, Newtons200 NMotorized up to 20 daN; manual up to 27 daNMotorized up to 20 daN; manual up to 30 daN178 maximum, motorized; 300 maximum, manual178 maximum, motorized; 300 maximum, manual178 maximum, motorized; 300 maximum, manualUp to 200 N200 maximumAdjustable maximum: 20 – 25 daNAdjustable maximum: 20 – 25 daNNeutons
Grid ratio6 to 1Grid ratio: 11:1 in both 2-D and DBT acquisitions2-D ratio 5:1 frequency 36l/cm – 3-D grid for DBT acquisition4:1; exclusive HTC grid4:1; exclusive HTC grid4:1 grid 31 lp/mm-low absorptionN/Ar5r50.209027778
BuckyN/A24 x 29 cm24 x 31 cmSealed; auto-retracting HTC gridSealed; auto-retracting HTC gridYesN/ALinear grid, motorizedLinear grid, motorizedNo Bucky, FFDM unit
Magnification deviceYes, 1.8x1.5 and 1.8 (2-D)1.5 and 1.8 (2-D)2-D: carbon fiber platform for 1.5x or 1.8x2-D: carbon fiber platform for 1.5x or 1.8x2-D: carbon fiber platform for 1.5x or 1.8x with fiagnostic imaging kit option1.8Diagnostic scanMag factors 1.6, 1.8Mag factors 1.6, 1.8, 2.0Yes
Stereotactic deviceAvailableN/SYes (2D)Optional stereotactic biopsy guidance system with tomosynthesis guidance optionOptional stereotactic biopsy guidance system with tomosynthesis guidance optionOptional stereotactic biopsy guidance system with tomosynthesis guidance optionYes using same digital detector areaN/ADigiGuideDigiGuideYes
Film ID systemN/ADigital detector, no filmDigital detector, no filmIncluded in the print functionIncluded in the print functionN/SN/AN/AN/ADICOM header
Label printerN/AN/SN/SN/AN/AN/SN/AN/AN/ANo
Power requirements200 V AC/208V AC/220 V AC (single pPhase)Input frequency: 50 Hz/60 Hz; input voltage: single-phase 200-240 V~; EATON UPS 5P650 650 VAInput frequency: 50Hz/60Hz; input voltage: single phase 200/208/220/240 V~, APC smart UPS 750 VAAWS: 100 120 200 220 230 240 VAC; gantry: 200 209 220 230 240 VAC nominal 10%, 5060 Hz 5% permanently wired; single- phase gantryAWS: 100 120 200 220 230 240 VAC; gantry: 200 209 220 230 240 VAC nominal 10%, 5060 Hz 5% permanently wired; single- phase gantryAWS: 100 120 200 220 230 240 VAC; gantry: 200 209 220 230 240 VAC nominal 10%, 5060 Hz 5% permanently wired; single-phase gantry8 kW output-240/220/208/200 VACNone208 VAC 50/60Hz – 240 VAC208 VAC 50/60Hz – 240 VAC208 V, 230 V, 227 V, +/-10% single phase, 208 V, 230 V 240V +/-10% two phase, 50/50 Hz+Hz
H x W x D, cm (inches)Gantry: 24.5 x 50 x 90 (62.4 x 127 x 228.5)Gantry: H 202 cm (80 in.) x W 72 cm (28 in.) x D 131 cm (52 in.); acquisition station: H 188 cm (74 in.) x W 77 cm (30 in.) x D 37 cm (15 in.)Gantry: H 193 cm (76in.) x W 70 cm (28 in.) x D 127 cm (50 in.); acquisition station: H 156 cm (61 in.) x W 71 cm (28 in.) x D 40 cm (16 in.); generator H 133 cm (52 in.) x W 64 cm (25 in.) x D 44 cm (17 in.)Gantry: 223 (max.) x 66 x 138 (87.8 x 26 x 54.25); acquisition station: 207 x 92.7 x 58.5 (81.5 x 36.5 x 23)Gantry: 223 (max.) x 66 x 138 (87.8 x 26 x 54.25); acquisition station: 207 x 92.7 x 58.5 (81.5 x 36.5 x 23)Gantry: 223 (max.) x 66 x 138 (87.8 x 26 x 54.25); acquisition station: 207 x 92.7 x 58.5 (81.5 x 36.5 x 23)144 x 93.5 x 21573.2 x 28 x 0540.4 x 39.1 x 29.7 in minimum40.4 x 39.1 x 29.7 in minimumN/S
Weight, kg (lb)370Gantry: 420 kg (926 lb); control station without monitors: 160 kg (353 lb)Acquisition station: 217 kg (478 lb); gantry: 420 kg (926 lb); generator: 160kg (353 lb)Gantry: 400 (882); AWS: 154 (340)Gantry: 400 (882); AWS: 154 (340)Gantry: 400 (882); AWS: 154 (340)33065 lbs421 lbs421 lbsExaminination stand and generator: 770 lbs; control desk with radiation sheild: 275 lbs
Optional accessoriesBiopsy paddle, rectangle spot compression paddle, flexAdditional 24 x 29 cm paddle; additional 19 x 23 cm sliding paddle; 24 x 29 cm flexibleOptions: 24 x 31 cm ergonomic paddle that conforms to the breast, 19 x 23 cm ergonomicAffirm biopsy guidance system; tomo guided biopsy software;Affirm biopsy guidance system; tomo guided biopsy software;Diagnostic imaging kit; Affirm biopsy guidance system; tomoStereotactic biopsy systems-Giotto Image 3D is pre-arrangedAccessory shelfN/SN/SN/S
paddles, stereotactice biopsy system compression paddle; 19 x 23 cm flexible and sliding compression paddle; 10 x 23 sliding implant/small breast compression paddle; square spot sliding compression paddle; round spot sliding paddle; 2-D localization 19 x 23 Swiss Cheese sliding compression paddle; 2-D localization 19 x 23 sliding standard compression paddle; 2-D crosshair device; X-ray protective shield; bar code reader; senoIris review station providing technologist easy access to prior images and reports in the exam roomsliding paddle that conforms to the breast, 1.5 and 1.8 magnification stands with dedicated paddles (19 x 23 cm, round spot, square spot), square spot sliding compression paddle, round spot sliding compression paddle, 2-D localizer with standard and round spot biopsy paddles, bar code reader, mobile mounting kit, stereotaxy add-on (2-D), senoIris Review station providing technologist easy access to prior images and reports in the exam roomC-view generated 2-D image software; I-View contrast enhanced 2-D software; SecurView DX diagnostic workstation; SecurView RT workstation; workstation cluster manager; advanced workflow manager; extra paddles; needle localization kit; CAD and Quantra Breast Density Analysis softwareC-view generated 2-D image software; I-View contrast enhanced 2-D software; SecurView DX diagnostic workstation; SecurView RT workstation; workstation cluster manager; advanced workflow manager; extra paddles; needle localization kit; CAD and Quantra Breast Density Analysis softwareguided biopsy software; C-view generated 2-D image software; I- View contrast enhanced 2-D software; SecurView DX diagnostic workstation; SecurView RT workstation; workstation cluster manager; advanced workflow manager; extra paddles; needle localization kit; CAD and Quantra breast density analysis softwarefor full field digital biopsy both in upright and prone position; the system can be transformed into a stereotactic prone biopsy system using same detector of the mammography unit and adding the MammoBed table
Warranty1 year1 year system warranty standard1 year system warranty standard1 year parts and labor; 2 year prorated manufacturers warranty on X-ray tube1 year parts and labor; 2 year prorated manufacturers warranty on X-ray tubeN/S1 year1 year1 yearN/S

updated 4.7.25