GE CT Scanner Comparison Chart
Model | Discovery CT750 HD | LightSpeed VCT | LightSpeed VCT Select | LightSpeed RT16 | BrightSpeed Elite | BrightSpeed Elite Select | Revolution Apex | Revolution CT | Revolution Fronteir Gen 2 | Revolution EVO Gen 3 | Revolution Maxima | Discovery RT Gen 2 | Optima CT540 | Optima CT520 | CardioGraphe | CT 580W | BRIVO CT325 | QX/I | Lightspeed Plus | Revolution Apex Elite | Revolution Apex Plus | Revolution Apex Select | Revolution Frontier Gen 3 | Revolution Ascend | Discovery CT590 RT | Optima CT580 RT |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Detectors, type | Gemstone scintillator | HiLight ceramic v-res | HiLight ceramic matrix 3 | HiLight ceramic matrix II | HiLight ceramic matrix II | HiLight ceramic matrix II | Gemstone scintillator | Gemstone scintillator | Gemstone scintillator | Clarity DAS | Clarity DAS | HiLight Ceramic Matrix II | HiLight Ceramic Matrix II | HiLight Ceramic Matrix II | High speed ceramic | Gemstone scintillator | Gemstone scintillator | Gemstone scintillator | HiLight ceramic scintillator | Gemstone scintillator | Gemstone scintillator | Gemstone scintillator | Gemstone scintillator | Clarity Detector and Data Acquisition system | Solid-state | Solid-state |
Slice count | 64 | 64 | 64 | 16 | 16 | 16 | 256 | 64 | 64 | 64 | 64 | 16 | 16 | 16 | 192 | 192 | 192 | 128 | 16 | 256 | 128 | 64 | 64 | 64 | 16 | 16 |
Gantry dimensions, H x W x D (cm) | 188.2 x 222.5 x 100.6 | 188.2 x 222.5 x 100.6 | 188.2 x 222.5 x 100.6 | 189 x 223 x 101 | 193 x 206 x 102 | 193 x 204 x 102 | 200.5 x 236.5 x 143.6 | 200.5 x 236.5 x 143.6 | 189.5 x 226.7 x 100.7 | 193 x 205 x 103.9 | 193 x 204 x 101 | 199.2 x 243.9 x 100.6 | 193 x 204 x 102 | 193 x 205 x 194.4 | 180 x 179 x 80 | 188.2 x 222.5 x 100.6 | 178.4 x 222.5 x 100.6 | 188.2 x 222.5 x 100.6 | 189 x 223 x 101 | 202.95 x 133.1 x 229.36 | 202.95 x 133.1 x 229.36 | 202.95 x 133.1 x 229.36 | 189.5 x 226.7 x 100.7 | 193 x 215 x 104 | 442 x 695 x 100.6 | 442 x 670.6 x 100 |
Gantry weight (kg) | 1850 | 1995 | 1995 | 1920 | 1770 | 1770 | 2,876 | 2,876 | 2,086 | 1,820 | 1,600 | 1,765 | 1,770 | 1,716 | 1,645 | 295 | 295 | 295 | 1,500 | 2798.7 | 2798.7 | 2798.7 | 2,086 | 1,805 | 1850 | 1820 |
Gantry aperture (cm) | 70 | 70 | 70 | 80 | 70 | 70 | 80 | 80 | 70 | 70 | 70 | 80 | 70 | 70 | 60 | 80 | 70 | 70 | 70 | 80 | 80 | 80 | 70 | 75 | 80 | 90 |
Heat storage, HU | 8,000,000 | 8,000,000 | 8,000,000 | 8,000,000 | 6,300,000 | 6,300,000 | 33,000,000 (equivalent value compared to conventional tube performance) | 6,800,000 | 8,000,000 | 7,000,000 (39,000,000 equivalent with ASiR-V) | 7,000,000 (39,000,000 equivalent with ASiR-V) | 8,000,000 | 6,300,000 (8,000,000 equivalent with ASiR | 3,500,000 (6,300,000 equivalent with ASiR | 3,050,000 | 3,050,000 | 3,050,000 | 2,700,000 | 6,300,000 | 33 | 33 | 33 | 8,000,000 | 7 MHU (50 MHU equivalent with ASIR-V, based on 72kW system) | 8 | 8 |
Tube Cooling | Oil/Air | Oil/Air | Oil/Air | Oil/Air | Oil/Air | Oil/Air | Oil/air | Oil/air | Oil/air | Oil/air | Oil/air | Oil/air | Oil/air | Oil/air | Oil/air | Oil/air | Oil/air | Air | Air | Oil/Air | Oil/Air | Oil/Air | Oil/Air | Oil/Air | Oil/Air | Oil/Air |
kW output | 100 | 72, 85 or 100 | 100 | 100 | 53.2 | 42 | 108 | 103 | 100 | 72 | 72 | 100 | 53.2 | 42 | 72kW, effective 139kW (equivalent value compared to conventional scanner geometry) | 100 | 100 | 120 | 120 | 108 | 108 | 108 | 100 | 55 (410 kW equivalent with ASIR-V); 72 (510 kW equivalent with ASIR-V) | 100 | 53 |
mA Range | 10 to 835 | 10 to 600, 700 or 800 | 10 to 800 | 10 to 800 | 10 to 440 | 10 to 350 | 10 to 1300 | 10 to 740 | 10 to 835 | 10 to 560 | 80, 100, 120, 140 | 10 to 800 | 10 to 440 | 10 to 350 | 50 -to 600, effective 100 to 1160 (equivalent value compared to conventional scanner geometry) | 10 to 835 | 200 | 10 to 835 | 40 to 1200 | 10 to 1,300 | 10 to 1,301 | 10 to 1,302 | 10 to 835 | 10 to 560 for 72kW system, 10 to 460 for 55kW system | 10 to 800 in 5 mA increments | 80, 100, 120, 140 |
Max scan volume, (cm) | 170 (200 option) | 170 (200 option) | 170 (200 option) | 170 (200 option) | 170 | 140 | 200 | 200 | 200 | 200 | 200 | 170 | 173 | 160 | 140 - axial | 170 | 170 | 170 | 160 | 50 x 50 x 50 | 70 x 70 x 70 | 60 x 60 x 60 | 80 x 80 x 80 | 75 x 75 x 75 | 170 | 170 |
Max load capacity with accuracy (lbs) | 227 | 227 | 227 | 227 | 227 | 205 | 227 | 227 | 227 | 227 | 227 | 227 | 227 | 205 | 227 | 295 | 200 | 425 | 450 | 227-306 | 227-306 | 227-306 | 227 | 227 | 650 | 500 |
Scan FOVs, (cm) | 25, 50 | 25, 50 | 25, 50 | 10 to 50 | 25, 50 | 25, 50 | 25, 50 | 25, 50 | 25, 50 | 25, 50 | 25, 50 | 25, 50 (10-80cm DFOV) | 25, 50 | 25, 50 | 25cm/16cm CTA, 35 cm Scout | 25, 50 | 30 | 50 | 40 | 25, 50, extended to 80 (MaxFOV, option) | 25, 50, extended to 80 (MaxFOV, option) | 25, 50, extended to 80 (MaxFOV, option) | 25,50 | 25, 50 extended to 75 cm with MaxFOV 2 | 50, 65 effective | 50, 65 effective |
Reconstruction matrices | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512, 1,024 X 1,024 | 512 x 512 | 512 x 512, 1,024 X 1,024 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 512 x 512 | 100 x100 | 1,024*1,024 | 1,024*1,025 | 1,024*1,026 | 512 x 512, 1,024 X 1,024 | 512 x 512, 1024X1024 | 512 | 512 |
Dose-modulation technique | 3D | 3D | 3D | 3D | 3D | 3D | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | N/A | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes |
Suggested room size (m2) | 34.2 | 34.2 | 34.2 | 28 | 18.2 | 18.2 | 28.6 | 28.6 | 28.6 | 25.2 | 25.2 | 22.6 | 23 | 22.1 | 15 | 34.2 | 34.2 | 34.2 | 30 | 28.6 | 28.6 | 28.6 | 28.6 | Over 18 for scan room | 28 (48 optimal) | 29 (48 optimal) |
kVp Range | 40 to 140 | 40 to 140 | 40 to 140 | 40 to 140 | 80 to 140 | 50 to 140 | 70 to 140 | 80 to 140 | 60 to 140 | 60 to 140 | 70 to 140 | 80 to 140 | 70 to 140 | 70 to 140 | 70 to 140 | 70 to 140 | 70 to 140 | 80 to 140 | 80 to 140 | 70 to 150 | 70 to 150 | 80 to 140 | 80 to 150 | 60 to 150 | 80,100,120,135 | 60 to 120 |