The Room Dimensions Required to Install a Linear Accelerator (LINAC)

This is a question we often field at ROS. It’s also not that easy to find the answer for various manufacturers in one place. We spent some time looking into the requirements for Varian, Elekta and Siemens. The following table shows the general room dimensions for the indicated makes and models.

 Linear Accelerator Optimal Room Height Optimal Room Width Optimal Room Length Base Frame Pit Dimensions (LxWXH)(ft.)
VARIAN
21/23 Series 9 ft. 20 ft. 25 ft. 13.75 x 5.5 x 12
6MV Series 9.5 ft. 20 ft. 22 ft. 10.2 x 5.5 x 1
Silhouette Series 9 ft. 20 ft. 22 ft. 12.2 x 5.5 x 1
or 21/23 Series VEO
ELEKTA
Synergy 10.5 ft. 19.6 ft. 20.7 ft. 4.6 x 5 x .77
Precise 10.5 ft. 19.7 ft. 20.7 ft. 4.6 x 5 x .8
SIEMENS
Oncor 9.5 ft. 20 ft. 20.5 ft. 11.1 x 5.25 x 1.1

Download our Linear Accelerator Buyer’s Guide for more information.

Linear Accelerator Buyers Guide

John’s role at Radiology Oncology Systems is to ensure that its customers receive the attention, service, and quality products they deserve. John does this by developing and managing the ROS team of world-class professionals who share the firm’s mission, and also by developing and nurturing the company’s growing network of vendors and partners.